Popular x64-bit Drivers Tags
- firmware x64-bit driver
- thermo fisher x64-bit driver
- analyzer x64-bit driver
Thermo Fisher 42i-HL Analyzer Firmware 02.02.00
Version 02.02.00 changes relative to version 02.00.06:
1. Fix sample mode digital output so it’s not active during purge mode.
2. Fix possible erroneous failure alarm and shutdown of permeation oven option that was introduced in version 02.00.06.
3. Fix lrec and srec CLINK commands to respond with only line feed character after each record and carriage return and line feed after the last record to correct response processing in iPort’s terminal window.
4. Fix softkey assignment screen to always show SAVING message for three seconds after the enter key is pressed.
5. Add support for new processor board revision 1.5T.
UPGRADE FIRMWARE
Below is a procedure for loading the firmware into FLASH memory. The firmware update file transfer process should take about 30 minutes at 57,600 baud. It is assumed that iPort is already talking to the instrument and the instrument window is currently open.
NOTE: DO NOT TURN OFF THE INSTRUMENT AT ANY TIME DURING THIS UPDATE If the instrument is turned off while burning the new image to the FLASH, it may require replacement of the CPU board, motherboard, I/O expansion board, and/or measurement interface board. To reduce this risk, make sure the instrument is running on clean and stable power before performing this update.
1. Close all instrument windows.
2. From the iPort menu, select Instrument > Update Firmware. Select TCP/IP or Serial, depending on the connection.
3. In the Update Instrument Firmware Program dialog box, enter the instrument ID (if using serial port) or the TCP/IP address (if using TCP/IP).
4. In the Open File dialog box, select the firmware update file, then click the Open button.
5. File transfer progress can be monitored by looking at the transferred blocks in the lower left corner of the iPort window as well as on the instrument's display.
6. Once the file transfer is complete, the instrument will automatically reboot. There may be some error messages regarding configuration and calibration files that are displayed, this is normal after a firmware update. At this time, the bootloader and application code in each of the low-level processors will be updated to the latest version.
7. To verify all updates were successful, go to the ALARMS menu and make sure the board status alarms at the bottom of the menu all show “OK”. If any board status alarms show “FAIL”, try rebooting the instrument and checking the ALARMS menu again. If they still show “FAIL”, contact technical service.
About Device Firmware:
Applying a newer firmware version than the one already installed on your unit can bring various enhancements, include workarounds for diverse problems encountered by the device, and improve or add newly developed features.
On the other hand, downgrading the firmware version can recover the device’s functionality in the event of a software update either turning up faulty or causing the unit’s overall performance to drop. However, bear in mind that applying an older firmware isn’t recommended and might not always be possible.
When it comes to the update steps, due to the large variety of devices and numerous ways for installing a new firmware, it is recommended that you carefully read and understand the upgrade steps, and consider applying a different firmware only when you have familiarized yourself with the update method.
As the installation process is most of the times quite risky, this step should be taken by users that have the ability to successfully complete the update; regular users may initiate it at their own risk. Moreover, it’s best that this task be performed in a steady power environment such as the one ensured by a UPS unit.
Therefore, if you consider applying this release, hit the download button, take into account all the aforementioned aspects, and update the device’s firmware. Also make sure you constantly check with our website to ensure that you don’t miss a single new release.
It is highly recommended to always use the most recent driver version available
Do not forget to check with our site as often as possible in order to stay updated on the latest drivers, software and games
Try to set a system restore point before installing a device driver
This will help if you installed a wrong driver
Problems can arise when your hardware device is too old or not supported any longer
Most popular Thermo Fisher 64-bit drivers
Members area
Top 64-bit Drivers
-
Samsung Electronics
System Firmware
10.0.12.67 for Windows
10 64-bit x64 driver -
Samsung Electronics
System Firmware
10.0.9.58 for Windows 10
version 1803 64-bit x64 driver -
Samsung Electronics
System Firmware
10.0.11.39 for Windows
10 Fall Creators 64-bit x64 driver -
Samsung Electronics
System Firmware
10.0.6.22 for Windows 10
64-bit x64 driver -
Wistron Dell System
Firmware 0.1.8.0 for
Windows 10 Creators
64-bit x64 driver -
Wistron Dell System
Firmware 0.1.24.0 for
Windows 10 Vibranium
64-bit x64 driver -
Wistron Dell System
Firmware 0.2.10.0 for
Windows 10 64-bit x64 driver -
HP ENVY 23-d001ej
TouchSmart Seagate HDD
Firmware HPD9/HPDB x64 driver -
Samsung System Firmware
10.0.2.42 for Windows 10
Fall Creators Update
64-bit x64 driver -
Samsung Electronics
System Firmware
10.0.12.67 for Windows
10 21H2 64-bit x64 driver -
Samsung Electronics
System Firmware
10.0.11.39 for Windows
10 21H2 64-bit x64 driver -
ASUS System Firmware
10.1.2.308 for Windows
10 Vibranium 64-bit x64 driver -
Dell NVMe Firmware
92.20.60.21 for Windows
10 21H2 64-bit x64 driver -
Wistron Dell System
Firmware 0.1.17.0 for
Windows 10 Vibranium
64-bit x64 driver -
ZOOM G3XN Multi-Effects
Pedals Firmware 2.00 x64 driver